Barnstead/Lab-Line MaxQ* 5000 Incubated & Refrigerated Floor Shakers
. Observe flask contents without opening the door through a large slanted viewing window
. Foot pedal opens the door automatically for hands-free operation
. A light bank is available on the low temperature shaker to mimic day and night conditions for plant and algae growth
Operation
. Shakers feature a triple eccentric drive to handle heavy workloads, provide uniform agitation and enable continuous 24-hour operation even at high speeds
. PID temperature controller precisely monitors and controls chamber temperature over complete range with ±0.1°C accuracy at 37°C
Analog Operating Systems
. DC motor and variable speed control from 40 to 400 rpm, controlled by a rotary dial
. Integrated tachometer monitors and displays speed in rpm to guarantee an accurate setting
. Easy-to-read LED digital display shows temperature
. Operate shaker in continuous mode or set timer from 1 to 60 minutes for time studies
Digital Operating Systems
. Solid state DC brushless motor, variable speed control from 15 to 500 rpm, and continuous and timed operation from 0.1 hour up to 99.9 hours or 0.1 minute to 99.9 minutes
. View speed, operating time and temperature simultaneously on 3 individual LED displays which helps minimize errors in establishing running conditions
. Visual/audible signals alert you to temperature deviations. Heat turns off if temperature deviates ±1°C of set point
. Shaker shut down and audible/visual alarms signal if unit operates ±10% of set speed preventing shaker from walking
. User adjustable speed calibration can be performed using a digital handheld tachometer where verification is required and protocols can be standardized
. Unbalanced load sensor stops platform motion when excess vibration is detected due to unbalanced load. Visual/audible alarm signal until condition is corrected
. Soft start feature eliminates splashing of vessel contents, wetting of flask closure, and helps prevent sudden starts and stops
. Set point retention retains parameters during power failure and restarts unit automatically after power is restored
. RS232 interface allows remote monitoring of speed, time and temperature
. Max. Load Capacity lbs (kg): 75 (34)
. Overall Dimensions in. (cm): 41 x 32 x 41 (104.1 x 81.3 x 104.1)
. Weight lbs (kg): 476 (215.9)